AUTOSPAN GLUCOSE

Test kit for in-vitro quantitative determination of Glucose in Human Serum/Plasma (Lyophilised Reagents )

  • Un-matched 12 months working reagent stability
  • Diluent provided for reconstitution
  • Black color bottle provided for storage of working reagent
  • Additional 400 mg/dL standard provided to validate high value samples
  • Programmable in End point as well as Kinetic mode
Method Glucose Oxidase - Peroxidase (GOD-POD)
Reagent System Lyophilized reagent & Diluent
Storage & Stability All reagents & standard are stable at 2~8℃ up to expiry, Diluent stable at room temperature (15~30℃) up to expiry
Working Reagent Stability After reconstitution, stable for 12 months at 2~8℃
Reagent Shelf Life 18 months


Assay Parameters

                                      
Method End PointFixed Rate Kinetic
Wavelength 505 nm505 nm
Blanking Reagent BlankPurified Water
Sample Volume 10 µl10 µl
Working Reagent Volume 1000 µl1000 µl
Incubation Time 10 Minutes--
Delay Time --60 seconds
Reading Time --60 seconds
Reaction Temperature 37℃37℃
Reaction Direction IncreasingIncreasing
Concentration of Standard 100 mg/dL (Calibration)
400 mg/dL (Linearity check)
100 mg/dL (Calibration)
400 mg/dL (Linearity check)
Linearity Up to 500 mg/dlUp to 700 mg/dl
Unit mg/dLmg/dL